Programming
EPM1000 User’s Guide
40
Terminates the current statistical batch and starts a new one.
Command Form
CHan STATistics START
Examples
Input
Response
ch stat start
<batch restarts>
Statistics Start
Set or query the trigger level of the EPM1000. Values accepted are 2
through 20.
Command Form
CHan TRIGgering LEVel <?|2..20>
Examples
Input
Response
ch trig lev 18
none
ch trig lev ?
CHAN A: TRIG: LEVEL 18
Trigger Level
Set or query the EPM1000 trigger source.
Command Form
CHan TRIGgering SouRCe
<?|INTernal|EXTernalPOSitive|EXTernalNEGative>
Examples
Input
Response
ch trig src extneg
none
ch trig src ?
CHAN A: TRIG: SOURCE EXTNEG
Trigger Source
Set or query the EPM1000 measurement units.
Command Form
CHan UNITS <?|VOLts|JOUles|WATts|HZ>
Examples
Input
Response
ch units jou
none
ch units ?
CHAN A: UNITS JOULES
Units