Operation
EPM1000 User’s Guide
25
Statistics Batch Restart Mode
The EPM1000 handles the end of a statistical batch in one of two
different ways. In Automatic Restart Mode, at the then of a batch, the
EPM1000 automatically begins collecting the next batch. The calculated
statistical values are displayed on screen until the new batch has been
collected. In Manual Restart Mode, at the end of a batch, the EPM1000
ignores additional incoming data and displays the finished batch's
statistical data until the EPM1000 is reset manually by pressing the
key or via the remote command:
CH STAT START
.
To change the Restart Mode, press . The LCD will display the
Statistics Batch Size and the Restart Mode. For example, if the
EPM1000 was set to a 1000-pulse Batch Size and Automatic Restart,
then the LCD would read
.
Use the and keys to toggle the restart mode from Automatic to
Manual.
Press to accept the selection or to cancel it.
System Settings
(Continued)
.
STAT
ENT
ZERO
Statistics Batch Size
Program the EPM1000 to collect a sample size from 2 to 9999 samples.
Press to change the Batch Size. The LCD will display the
statistics batch size and the restart mode. For example, a batch size of
6126 with manual restart appears as
.
Use the Numeric Entry Keys to change the value of the batch size.
Press to accept the new value or to cancel the action.
.
STAT
ENT
ZERO
Wavelength Compensation (for Thermopile only)
Sometimes a probe calibrated for one laser wavelength must be used to
measure laser energy of another wavelength. Because of differing
absorptive characteristics, the results are often incorrect.
The EPM1000 can be programmed to compensate for such a change in
wavelength.
To activate Wavelength Compensation, press .
To change the compensation value, press . Use the numeric
entry keys to change the decimal value and the and keys to
select µm or nm. Press at any time while entering data to restore
the initial value. Press to complete the action.
To deactivate Wavelength Compensation, press .
ENT
.
Hz
.
Hz
ZERO
ENT
ENT
.
Hz
ZERO
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com