Programming
EPM1000 User’s Guide
36
Command Form
OUTputFORMat <?|BINary|ASCII|ASCII+>
Examples
Input
Response
out form bin
none
out form ?
OUTPUT: FORMAT BINARY
Output Format
(Continued)
ASCII+
A commonly used form of data transmission, most easily readable. Data
is indicated using scientific notation and units are included. For example,
0.687E+00V
.
When multiple values are transmitted (such as statistical data), the values
are separated by a comma. To avoid confusion, data headers should be
utilized when transmitting multiple values (see Programming --
Commands -- Statistics Send). An example of multiple values with
headers is:
AVG=0.950E+00V,DEV=0.001E+00V
.
Probe Type
Query the EPM1000 to determine the type of probe attached. In the
case of a PowerMax
®
probe, the EPM1000 will also return the head code
in hexadecimal notation.
Command Form
CHan TYPE ?
Example
Input
Response
ch type ?
CHAN A: TYPE PYRO
Range, Current
Set or query the EPM1000's current range. Example range arguments
include
300mv
,
30mj
,
3j
, etc.
Command Form
CHan RANGE <?|
range argument>
Example
Input
Response
ch range 10V
none
ch range ?
CHAN A: RANGE 10V