78
OPERATOR'S MANUAL
EN
ERRORS (CATEGORY E)
CODE
MESSAGE
ON
THE
DISPLAY
DESCRIPTION
EXIT
E000
BLACKOUT
Black-Out.
3
E001
OVERVOLTAGE
Overvoltage.
3
E004
NETWORK FREQUENCY
ERROR
Network frequency reading problem
3
E011
LID OPEN!
Attempt to start a cycle with lid open.
2
E552
WAIT!
Attempt to start a cycle with chamber
drainage in progress after full chamber
alarm
2
E070
MIN SALT LEVEL
Signal of salt float minimum level. It
locks the machine after a number of
cycles depending on set water
hardness
2
E071
START CYCLE WITH
REGENERATION
Signal to start a cycle with resin
regeneration (Resin regeneration,
Disinf. 90°C, Washing, D1 Custom, D2
Custom and W1 Custom)
2
E800
MIN DETERGENT LEVEL
Detergent fluid level under the
minimum level
2
E900
NO SET PW US
Power read by US board memory other
than requested value
3
E901
POWER ABOVE / UNDER
THRESHOLD
Power detected by US board other than
set value
3
E957
UPDATING US IN
PROGRESS
Ultrasound update in progress
2
E999
MANUAL INTERRUPTION
Manually interrupting the cycle
3
1 = OK (warning) + chamber emptying
2 = OK (warning)
3 = cycle interr ok + door reset 3