EN
OPERATOR'S MANUAL
41
After having inserted the load in the thermo-disinfector basket (following the precautions
indicated in 'Preparing the material' chapter) select the desired cycle by pressing the button:
Start the cycle by pressing the button corresponding to THERMAL DISINFECTION.
The cycle counter appears in the upper left corner.
1. total cycle time
2. disinfection time
3. A0 (foreseen for the selected
cycle)
4. Process nominal temperature
NOTE
A0 THERMAL DISINFECTION PARAMETER
A0 PARAMETER (INTRODUCED BY STANDARD EN 15883) ALLOWS TO
ASSOCIATE A NUMERIC VALUE TO THE THERMAL DISINFECTION
LEVEL REACHED.
A0 IS CALCULATED, STARTING FROM THE OVERCOMING OF 65°C
DURING HEATING PHASE AND SHOWN STARTING FROM THERMAL
DISINFECTION PHASE, BY STEPS OF 50 UNITS.
WARNING
AT THE END OF THE CYCLE IT IS ESSENTIAL TO VISUALLY INSPECT
THE OBJECTS TO CHECK THEIR PROPER WASHING.
WARNING
THE FIRST TIME THE DEVICE IS USED AFTER THE INSTALLATION AND
AFTER A PERIOD OF INACTIVITY EXCEEDING 24 HOURS IT IS
NECESSARY TO CARRY OUT AN EMPTY 90 CYCLE (DISINFECTION 90°)
BEFORE USING THE DEVICE.