System Two LED Installation and Service Manual (1003385 Rev. C)
Page 56
Section 6
Section 6: General Maintenance
General Safety
Electric shock hazard! The Arm
System can carry an electric
current and must be treated with
the utmost care during cleaning
and disinfection
Only facility authorized
maintenance personnel should
troubleshoot and/or repair the
Lighting System.
Troubleshooting by unauthorized
personnel could result in
personal injury and/or property
damage.
No modification of this equipment
is allowed. Do not modify this
equipment without authorization
of the manufacturer. If this
equipment is modified,
appropriate inspection and
testing must be conducted to
ensure continued safe use of the
equipment.
Isolation from Supply Mains- The
isolation shall be provided by the
facility. Appropriate Breakers
with simultaneous action to all
poles shall be made available.
The articulating arm is spring
loaded. When removing the
head/yoke ensure that the arm
lock pin has been installed.
Failure to do so could result in
personal injury and/or property
damage.
The articulating arm is spring
loaded. Never remove the arm
lock pin until the head/yoke has
been installed. Failure to do so
could result in personal injury
and/or property damage.
Follow the product
manufacturer’s instructions.
Failure to do so could result in
personal injury and/or property
damage.
If the unit fails any part of the preventive
maintenance functional checks, repair the unit
before use on any patient. Failure to do so could
result in personal injury and/or property damage.
Do not use harsh cleaners, solvents, or
detergents. Failure to do so could result in
equipment damage.
Do not use silicone based lubricants. Equipment
damage could occur.
The front lens is supplied with a protective hard
coat to resist scratching. Never use abrasive
cleaners on the front lens. Failure to do so could
result in equipment damage.
Turn off main power before any repairs are
started. Failure to do so could result in personal
injury and/or property damage.
Do not pinch any wires during installation.
Pinched wires can cause an electrical shock
hazard, resulting in personal injury and/or
property damage.
Use only Medical Illumination fuses P/N 0001475
(250 VAC/1.6A slow-blow, Break Capacity: 150%,
1.6A @ 60 minutes min. 1000%, 1.6A @ 20ms min
to 300ms max.) if replacement is necessary.
Failure to do so could result in personal injury
and/or property damage.
Do not expose the unit to excessive moisture.
Failure to do so could result in personal injury
and/or property damage.
Do not rest articles or liquids on top of the System
Two LED Surgical Lighting System. Spilled liquids
will damage the light head and arm assemblies
causing an electric shock hazard.
Use only a White Solid Lubricant Paste Grease
(white grease) in lubricating securing segments.