![MJK MagFlux Installation And User Manual Download Page 52](http://html1.mh-extra.com/html/mjk/magflux/magflux_installation-and-user-manual_1812930052.webp)
52
GB MagFlux Manual 0812
Electromagnetic Flow Meter
Default
When "Default" is selected, the following settings
are not
affected and
remain as chosen by the user:
Sensor Name
•
Device ID
•
Flow Direction
•
DN, size
•
Calibration
•
Calibration Code
•
The in- and outputs are
not
activated from the factory. Activate the in- and
outputs by configuring the functions with for example the default values
shown in the following tables:
MagFlux Default DI/DO Settings
•
MagFlux Default 20 mA Settings
•
MagFlux Default DI/DO Settings
Digital Output 1 (opto) - DO1
Digital Output 2 (mech.) - DO2
Digital Input - DI
Counter R Totalizer Forward
Counter R Totalizer Forward
RESET R Totalizer Forward
Counter R Totalizer Reverse
Counter R Totalizer Reverse
RESET R Totalizer Reverse
Batch Counter 1
Batch Counter 1
RESET R Totalizer Sum
Batch Counter 2
Batch Counter 2
Start / Pause Batch counter 1
Low Flow
Low Flow
Start / Pause Batch counter 2
High Flow
High Flow
Flow Direction
Flow Direction
Empty Pipe
Empty Pipe
Note: The same function can not be on assigned to both relays.
Summary of Contents for MagFlux
Page 6: ...6 GB MagFlux Manual 0812 Electromagnetic Flow Meter This page intentionally left blank...
Page 12: ...12 GB MagFlux Manual 0812 Electromagnetic Flow Meter This page intentionally left blank...
Page 38: ...38 GB MagFlux Manual 0812 Electromagnetic Flow Meter This page intentionally left blank...
Page 106: ...106 GB MagFlux Manual 0812 Electromagnetic Flow Meter This page intentionally left blank...
Page 144: ...144 GB MagFlux Manual 0812 Electromagnetic Flow Meter This page intentionally left blank...