Ingenio® EZporator® Electroporation System
Product User Manual
| Instructions for MIR 51000
Set
-
up Instructions
Installation
Place the EZporator® Pulse Generator on a laboratory bench or work table. Be sure
to allow 3 to 5 cm of clearance on all sides of the module for proper cooling. When
selecting a workspace, choose a stable, dry location with an easily accessible power
outlet. NOTE: It is normal for the EZporator® Pulse Generator module to be slightly
warmer than its operating environment.
Connect the power cord to the EZporator® Pulse Generator module AC Input on the
back panel, then insert the power cord plug into an appropriate electrical outlet (i.e.
110
-
240 VAC).
To connect the EZporator® Cuvette Chamber to the Pulse Generator, insert the color
coded plugs attached to the Cuvette Chamber into the HV Output located on front
panel of the EZporator® Pulse Generator.
Initializing
1.
Turn the EZporator® Pulse Generator on by pushing the green power button
located at the back of the instrument (see page 19). NOTE: Push the power
button once to turn on. Push again to turn off. Do not hold the power button.
2.
The EZporator® will go through a series of self
-
test algorithms to test generator
functionality. The display will flash
“
EZporator®
”
and the software version (e.g.
V1.01) prior to attaining a ready status.
3.
When the instrument is initialized for the first time, the display will read
“
READY LV 400V.
”
Instrument Features and Controls:
Cuvette Chamber
High Voltage
Output
LCD
Pulse Button
Voltage Adjust