System Components and Features
The EZporator® Electroporation System was designed for mammalian and insect cell
transfection with Ingenio® Electroporation Kits and Solution.
System features include:
•
Digital Liquid Crystal Display (LCD) user interface
•
Pulse Generator to deliver exponential decay pulse at user
-
determined voltage.
NOTE: In exponential decay pulse conditions, a set voltage is released from the
capacitor and decays exponentially.
•
Cuvette Chamber compatible with 0.1 cm, 0.2 cm or 0.4 cm cuvettes
•
Power cords with Type B, F and G electrical outlet plugs
•
Low Voltage (LV) mode
-
Voltage range of 20 to 400 V with 2 V resolution
-
150
Ω
internal resistance and 1,050 µF capacitance
•
High Voltage (HV) mode
-
Voltage range of 30 to 2,500 V with 10 V resolution
-
150
Ω
internal resistance and 36 µF capacitance
•
Streamlined experimental approach: (1) Initialize instrument, (2) Set voltage
and mode with the Voltage Adjust knob and (3) Deliver electroporation pulse
by pressing the Pulse button.
•
Dual monitoring and display of peak output voltage and time constant
System Components:
Upon receipt, carefully open the box containing the EZporator® Electroporation
System and verify that the box contains the following:
•
EZporator® Pulse Generator (1)
•
EZporator® Cuvette Chamber (1)
•
Power cords (3)