Step 2 Clean the Handpiece
Once the bioTip has been removed, the “Clean Handpiece” Screen will appear as shown in
Figure 28. Clean any excess water-based lubricant that may be remaining on the Handpiece
using isopropyl alcohol.
Figure 28: “Clean Handpiece” Screen
Step 3 Place the Handpiece in the Console holster
Place the Handpiece back in the holster in the “down” position as shown in Figure 7.
Step 4 Review “Treatment Summary” Screen
A “Treatment Summary” Screen will appear as shown in the example in Figure 29. The screen
summarizes the number of completed placements at each microwave energy level and the
guided treatment template size, for each axilla.
Figure 29: “Treatment Summary” Screen (example with both axilla treated in Guided Mode)
Step 5 Place an Ice Pack on the Axilla Treatment Area
After the procedure, clean the patient’s skin and place an icepack on the treatment area.
9.7.
OPERATION – TOUCH-UP MODE
Follow the instructions in Section 9.3 for selecting Touch-up Treatment Mode, applying a
Touch-up Template, and administering anesthesia. This includes any special safety instructions
for administering anesthesia for touch-up placements. Then follow the instructions in sections
9.2 and 9.4 for selecting the appropriate microwave energy level.