ALWAYS wait a minimum of 10 minutes after administering anesthesia to initiate
treatment. NEVER initiate treatment if more than 60 minutes have elapsed since
administering anesthesia.
•
DO NOT use the miraDry System unless the axilla size has been measured and
selected appropriately on the Touchscreen GUI (unless you are operating in Touch-
up Mode). The axilla size selection determines the recommendations for anesthesia
administration.
•
ALWAYS place ice packs, wrapped in a towel on the patient’s treated skin after the
procedure.
•
DO NOT use lubricants other than the recommended lubricants (KY® jelly or KY®
liquid). Use of other substances such as ultrasound gel or IPL laser coupling gel can
reduce the effectiveness of the surface cooling, leading to blisters or burns.
•
DO NOT use the miraDry System unless an appropriate amount of skin lubricant is
applied to the axilla treatment area.
•
DO NOT use the miraDry System unless the skin surface is wiped clean using
antiseptics before treating.
8.4.
CLEANING WARNINGS AND PRECAUTIONS
•
ALWAYS turn the miraDry Console
OFF and disconnect the AC Power Supply Cord from
the AC power supply before cleaning.
•
ALWAYS clean, disinfect, and maintain the electrical units in accordance with the
operating instructions.
•
ALWAYS clean and remove any dust, lint or particulate materials from the air intake
screen filter located underneath the miraDry Console. Use a vacuum cleaner and/or
water. Make sure to remove the filter from the miraDry Console before cleaning. This
filter is attached to the bottom of the Console using magnetic attachment and is
easily removed. Thoroughly dry the filter before reattaching it to the Console.
8.5.
MAINTENANCE WARNINGS AND PRECAUTIONS
•
DO NOT remove the covers of any of the miraDry System components. There are no
user-serviceable parts inside. Opening the cover of any component may result in
electrical shock or thermal burns.
•
DO NOT attempt to repair or alter any of the miraDry System components.
9.
OPERATING INSTRUCTIONS
When you are ready to unpack and set up the miraDry System, contact a Miramar Labs
Representative for assistance. It is important that a Miramar Labs Representative be
present and set the System up for you. DO NOT attempt to set up the System without the
help of a Miramar Labs Representative.
9.1.
OPERATION – PRE PATIENT SETUP
Step 1 Attach the Handpiece to the Console
Attach the miraDry Handpiece to the miraDry Console as shown in Figures 2a, b and c. Use
one hand to stabilize the Handpiece Cable Connector and the other to turn the black Handpiece
Collar. Screw the Handpiece Collar clockwise onto the male connector threads on the Console
until the red ring is fully concealed.