It may take up to 20
minutes to complete treatment on one axilla. It is important to maintain a
comfortable posture and proper technique while administering treatment to minimize neck, back
and hand strain.
Skipping a Template Placement
You can skip a Template placement or return to redo a Template placement in Guided Mode.
Simply press the wrench button (
) at the top right corner of the Guided Treatment Mode
Screens. This lets you change the status of a Template placement from “Pending” to
“Completed”, or “Completed” to “Pending”. Skipping a Template may be useful when you want
to avoid treating a particular area due to a pre-existing condition, like a scar or large mole. It is
also useful when you want to redo a placement that was not satisfactory.
Touching the wrench button (
) button shown in Figure 24 will remove the Handpiece image
from the Touchscreen GUI and highlight each placement flag (A, B or C) with the same blue
outline found on other buttons. Touching a placement flag will switch its status back and forth
between “Pending” (black) and “Completed” (grey). Touching the wrench button again shown in
Figure 25 will remove the blue outlines from the placement flags and let you resume treatment.
Figure 24: Active Wrench Button, Available
Figure 25: Wrench Button Not Active, Mode Not Turned On
9.5.
OPERATION – STOPPING A TREATMENT
There are different ways to stop a treatment while it is in progress:
•
Stop Treatment in the Middle of a Placement
There are two ways to stop microwave energy delivery while administering treatment at a
particular placement. The effect of this action is to end the treatment cycle at that one
particular target site but be free to proceed with the next placement/treatment.
1.
Press down on the Console footswitch (optional) while the miraDry Handpiece is
administering treatment.
2.
Press the Handpiece handswitch while it is administering treatment.
Both methods place the Handpiece in post-cool mode.
Continue to keep the Handpiece
in contact with the skin through the entire post-cool period
.