LB0261 Rev A 01/2018
© Miramar Labs, Inc. 2018, Confidential Information of Miramar Labs, Inc.
Page 29
Figure 21: Treating the 1A position
Firmly place the Handpiece on patient’s skin as shown in Figure 22 to start the first treatment
cycle. Make sure the full perimeter of the flexible skirt of the miraDry bioTip is in contact with
the patient’s skin.
Figure 22: Firmly place Handpiece on patient's skin
Step 4 Start Treatment
Initiate treatment by either pressing the handswitch on the miraDry Handpiece or pressing the
footswitch (optional) attached to the miraDry Console.
Using suction created from the Console vacuum, skin tissue is first drawn into the bioTip
acquisition chamber and then placed against the cooling plate. While the skin is in the chamber,
microwave energy is delivered to the target site for the selected duration time. Following
microwave energy delivery, the skin is cooled and then released from the chamber. An audible
“treatment” tone will sound during this entire treatment cycle.
If you press the handswitch or footswitch a second time during the treatment cycle, you will
terminate the treatment cycle. If termination occurred during the short, initial skin acquisition
phase, the System will go back into ready mode. If termination occurred during the microwave
energy delivery phase, the System will skip directly into the post-cool phase before going back
into ready mode.
Handswitch