Figure 5: “Add Water” Popup Prompt
When the W01 add water popup appears, remove the plastic cap at the water fill location by
unscrewing it counter-clockwise. Pour the water over the top of the filter to fill the console, and
the water will flow through the membrane.
Fill the water until the water level is all of the way
up the threads above the perforated metal shielded membrane filter (Figure 6), even
though a beep will sound and the “Add Water” popup prompt will close well before the
water reaches this level.
Figure 6: Perforated Metal Shielded Membrane Filter
If the miraDry Handpiece has just been attached for the first time, you will be prompted to add
more water with a W08 popup message. It will take approximately 40 ml of deionized or distilled
water to fill a completely empty Handpiece.
Step 5 Place the System for Patient Treatment
Position the System close to the patient and make sure none of the air vents located at the
bottom or rear of the Console are obstructed.
Step 6 Position the Handpiece in the Holster to Start the System Self-Test
The System performs a self-test when power is turned ON to ensure that all functions are
working as expected.
Place the Handpiece in the holster and lower it to the “down” position as shown in Figure 7. The
Handpiece must be in the “down” position to complete the self-test.
Perforated Metal Shielded
Membrane Filter