Figure 20: Change Template Screen for second axilla.
If power is lost once in the treatment screen and energy has been delivered for a placement,
the system will maintain its position in the treatment sequence. However, during the power-
cycle recovery, an appropriate pop-up screen will first appear indicating the side and template
size which was last used, without the Change Template button.
Step 2 Lubricate the Axilla
Place a small amount of KY® jelly or KY® liquid (water-based lubricant) on the patient’s skin.
Use approximately 1-4 ml of lubricant on each axilla, depending on the size of the axilla and
dryness of skin.
DO NOT use the miraDry System unless an appropriate amount of skin lubricant is
used.
Step 3 Position the Handpiece on the Axilla
Place the miraDry Handpiece in the appropriate position on the patient’s skin. For Guided
Treatment Mode, the Touchscreen GUI schematically shows the first treatment position as 1A.
This should match the treatment position on the patient’s skin that has been created from the
skin Template.
A treatment cycle begins and ends with microwave energy delivery to a particular target site.
Triangles on the Template grid indicate the various treatment target sites. The green triangle
shown in Figure 21 indicates the target site that is ready to be treated. While microwave energy
is being delivered to the target site, the triangle remains green. The triangle turns grey when
microwave energy delivery stops to this target site, signaling the end of the treatment cycle.
The next triangle in sequence will turn green to indicate the target site of the next treatment
cycle.