Spirolab
Rev.1.8
Page 31 of 60
This section allows to choose the
unit of measure to show the PEF
value: L/sec or L/min
In the "Incentive" section, you can
configure the paediatric incentive
to
improve
spirometric
performance.
The incentive may be enabled or
disabled; should it be enabled, by
selecting "ON" on the left, you
must enter the age threshold up to
which the incentive is displayed;
use the numeric keypad on the
bottom to set this value.
Calibration
For calibration, see paragraph 2.5.1 below.
Oximetry
In this section, you can configure
all alarms related to an oximetry
test. By setting the desired alarms
to "ON", you will activate a box
where you can set the relevant
threshold value using the
numeric keypad.
WARNING
If the maximum value of a parameter is less than or equal to the minimum value,
the setting process will not proceed. A beep will sound and the device will
automatically return to the screen for setting the minimum value.
References
This section allows you to set the predicted value used between:
Adults
Children
ERS
Knudson
Knudson