Spirolab
Rev.1.8
Page 54 of 60
•
If no beeps are emitted during this movement, clean the turbine.
4.2
Cleaning the oximetry sensor
The reusable oximetry sensor must be cleaned at every change of the patient, i.e. before
applying it to a new patient.
Clean the sensor with a damp cloth containing water or a delicate soap solution. To
disinfect the sensor, rub it with isopropyl alcohol. After cleaning, allow the sensor to fully
dry.
Do not use abrasive or caustic agents to clean the sensor.
WARNING
Do not sterilise by means of radiation, steam or ethylene oxide.
Always disconnect the sensor before cleaning or disinfection.
Spirolab
sensors are latex free.
4.3
Recharging the battery pack
WARNING
To charge the battery pack, only use the charger supplied by the manufacturer.
When the
spirolab
device is turned on, an icon shows the condition of the battery in the
top right hand corner:
The maximum charge level is reached when the symbol is completely green.
If only part of the symbol is green or if the device doesn't turn on, you will need to recharge
the battery pack as follows:
•
Connect the battery charger to both the connector and
the mains supply. A blue LED will light up in the lower
right-hand corner.
•
When charging is completed, the LED will turn green.