![Minebea CSD-581-15 Instruction Manual Download Page 52](http://html.mh-extra.com/html/minebea/csd-581-15/csd-581-15_instruction-manual_1801110052.webp)
32
Procedures
1
Press the “FUNC.” key for about one second.
The load display will show “FUNC”.
2
Press the “A/Z
A
” key three times.
It will make the load display proceed as “FUNC”
→
“CCAL”
→
“ACAL”
→
“LCAL”.
3
Press the “ENTER” key.
“LCAL” mode can be entered, and the load
display shows “D
−
01” and lights on and off.
When the calibration has been already
completed, the set value of minimum scale
registered at that time will be displayed.
Set the minimum scale with “A/Z OFF
Y
” key.
Set values for the minimum scale are 4 (four)
kinds as follows:
1, 2, 5, 10
“A/Z OFF
Y
” key
:
Set value Increment key
4
Press the “ENTER” key.
The load display will show “DISP”.