![Minebea CSD-581-15 Instruction Manual Download Page 109](http://html.mh-extra.com/html/minebea/csd-581-15/csd-581-15_instruction-manual_1801110109.webp)
89
Set the connection range to DC
・
V for the measuring instrument such as tester and
so on.
①
Remove the connecting cable from the terminals on the strain gage applied transducer.
②
Measure the voltage between terminal A and terminal C.
①
Re
−
connect the connecting cable with the terminals on the strain gage applied
transducer. (Refer to the chapter 4.)
②
Referring to 7
−
15, set the load display to the monitoring condition for strain
gage applied transducer.
Monitor display exceeds
the value from
−
0.1 mV/V
to 2.4 mV/V.
7
Inform
us
about
the
contents
of
failure
and
situation
at
site
in
details.
Short between
terminal B and D.
Start of
measurement
Same condition
Make re
−
calibration
according to the procedures
in the chapter 5.
YES
Following causes can be considered.
①
Input of strain gage applied
transducer exceeds the value from
−
0.1 mV/V to 2.4 mV/V.
②
Strain gage applied transducer
has destroyed due to overload and
so on.
③
Signal route (signal wire) is cut
off for the strain gage applied
transducer.
YES
NO
YES
NO
NO
NO
Monitor display exceeds
the value from
−
0.1 mV/V
to 2.4 mV/V.
YES
Contact
us
.
The voltage between A and C
is just like the setting of SW1.