17
Internal terminal connections for SB−310
To CSD−581−15
Suitable crimp−type
terminal lugs
・
Y type
RD
Short bars
Output
Port
Input port No.1
Input port No.2
Input Port
No.4
or
WH
YE
YE
GN
BL
WHGN BLYE
YE
YE
BL
GN
WH
WH
RD
GN BL
BL
GN
WH
φ
A
φ
B
RD
RD
RD
Input port No.3
A
:
3.2 mm or less
B
:
7 mm or less
・
Round type
BK
OR
RD
B
A
4−3−2. Connection with external control input
Connection with external control input of “ZERO”, “A/Z”, “A/Z OFF”, “HOLD” , “LOCK”
“SEL 1” and “SEL 2” should be made by using contact or open collector as the below figure
indicates. Refer to the paragraph 7−1. External control input signal, open collector output
signal for each input functions.
or
10
・
・
・
・
・
・
27
36
1.19
33
32
31
30
29
28
Connector
plug
57−30360
To connector plug case
Shield
Warning
●
Connections with external control input should be made securely as
indicating in the above figure. If neglected, it may cause an
unexpected malfunction to the instrument.
●
For the connection with external control input, be sure to apply
shielded cable, and shielded cable should be connected with the
connector plug case. If not connected, it may cause malfunction due
to the effect from external noises and so on.
CN202
DIP202 DIP203