12
※
1 When tension is applied with the application of tension type or
universal(compression/tension) type of load cell, and display of “+”
direction is required, connect “Green” with Terminal No.9 at
terminals 1, and “Blue” with Terminal No.8 at terminals 1
individually. As there is a case which standard wiring color is
different, please confirm the inspection data sheet of the load cell
being used.
※
2 When the total length of CAB−502 specifies more than 30 m, the
accuracy may be out of warranty because the resistance of cable
makes the input voltage of the instrument decreased.
※
3 When the total length of CAB−501 specifies more than 100 m, the
accuracy may be out of warranty, because the resistance of cable
makes the remote sensing function of the instrument deteriorated.
3
Connection with 2 to 4 points of Load cells and Summing type Junction box with balance
adjustment applied
(
B−307
)
and Digital indicator CSD−581−15
i) In case of application of CAB−502
(
4 cores cable)
※
2 (Length of CAB−502 should be
30 m totally.)
B−307
Junction box
CAB−502
Enlarged figure for connecting section
WH
T
※
1G
R
N
※
1B
L
U
Sh
ie
ld
RED
Attached short bars
Attached short bars
RED
WH
T
※
1B
L
U
Sh
ield
※
1G
R
N
CN202
DIP202 DIP203