36
6-11. Ending calibration
Write the data in the following table to store your provisionally registered calibration data, such as zero and
span point data, and return to measurement mode.
Writing data other than [0] to the general data area causes an out-of-range error.
㻌
(Error code: 99, auxiliary error code: 1)
Command no.
(RWwnE): 4 stations shared
(RWwn6): 2 stations shared
General data area
(RWwnC~RWwnD): 4 stations shared
(RWwn4~RWwn5): 2 stations shared
9099
0
Unexecutable error
The following codes are set if performed in modes other than calibration mode or if not set according to the
flowchart at the beginning of this chapter:
Status
Error code
Auxiliary error code
Unexecutable error
99
2
z
All LEDs on the indicator are illuminated when calibration mode is exited before entering
measurement mode.
z
Because this fully illuminated state is not measurement mode, error code 3 and auxiliary error code
99 occur.
6-12. Forced termination of calibration
Write the data in the following table during the calibration procedure to stop calibration and return to
measurement mode without storing your provisionally registered calibration data.
Writing data other than [0] to the general data area causes an out-of-range error.
㻌
(Error code: 99, auxiliary error code: 1)
Command no.
(RWwnE): 4 stations shared
(RWwn6): 2 stations shared
General data area
(RWwnC~RWwnD): 4 stations shared
(RWwn4~RWwn5): 2 stations shared
9000
0
Unexecutable error
The following codes are set if performed in modes other than calibration mode or if not set according to the
flowchart at the beginning of this chapter:
Status
Error code
Auxiliary error code
Unexecutable error
99
2
z
All LEDs on the indicator are illuminated when calibration mode is exited before entering
measurement mode.
z
Because this fully illuminated state is not measurement mode, error code 3 and auxiliary error code
99 occur.