6-10. Entering load cell output voltage values to register span point
Write the data in the following table to enter a numerical value for registering the span point. Values can be
registered in 0.00001mV/v increments.
Command no.
(RWwnE): 4 stations shared
(RWwn6): 2 stations shared
General data area
(RWwnC~RWwnD): 4 stations shared
(RWwn4~RWwn5): 2 stations shared
9011
Data range: 1~310000
(0.00001 mV/V~3.1 mV/V)
SP-L error
The following codes are set if the load cell output value set is less than 0 mV/V and not within the span
adjustment range (span point load cell output value -
zero point load cell output value
≤
0.0 mV/V):
Status
Error code
Auxiliary error code
SP-L error
4
2
SP-H error
The following codes are set if the load cell output value set exceeds 3.1 mV/V and the span adjustment
range:
Status
Error code
Auxiliary error code
SP-H error
4
3
Unexecutable error
The following codes are set if performed in modes other than calibration mode or if not set according to the
flowchart at the beginning of this chapter:
Status
Error code
Auxiliary error code
Unexecutable error
99
2