— CAL
Initiates a two-point calibration (zero and max.
load)
— LIN
Initiates a linearization (zero, half weighing
range and max. load)
— GEO
The adjustment of the calibration based on the
current location is carried out using the
geographic adjustment factor (GEO). (Settings
from 0 … 31, default value = 12)
— END
Exit menu level
5.4
Calibration and adjustment
The scale can be calibrated and adjusted in this menu.
Initial calibration
If the scale is being put into operation for the first time, calibration is recommended in
order to ensure precise weighing results. Before the calibration, ensure that the
appropriate calibration weights are available.
Adjust the GEO setting according to the location (see Table
).
CAL.ADJ
5.4.1
[CAL] calibration
Call up the menu mode:
Release the key.
1.
Press and hold
Menu
M
+
until [M.E.N.U] is displayed.
The first menu item [APPLIC] (application) is displayed.
2.
Press the
Unit
key in order to go to the next menu item, or press the
Mode
F
key to go
to the previous menu item.
3.
Change the menu item until [CAL.ADJ] is displayed.
4. Press the
Off
(Yes) key to go to the sub-menu item.
The sub-menu item [CAL] (calibration) is displayed.
5 Operating
Floor scale Puro®
Minebea Intec
EN-41