3
Device installation
3.1
Package Contents
-
Scale
-
Sub-platform
-
Safety Instructions and QR code with access to the detailed documentation
-
Pan
-
USB power supply and cord
3.2
Requirements on location
-
Set up the device on a stable, even surface.
-
Position the device so that the power plug is freely accessible and the power cord is
not an obstacle or trip hazard.
-
Avoid placing the device in close proximity to a heater or otherwise exposing the
scale to heat or direct sunlight.
-
Do not expose the device to excessive temperature flutuations.
-
Protect the device from drafts that come from open windows or doors.
-
Avoid exposing the device to extreme vibrations during weighing.
-
Protect the device from aggressive chemical vapors.
-
Do not expose the device to extreme moisture over long periods.
NOTICE
Acclimatizing the device
Condensation can form on the surfaces of a cold device when it is brought into a
substantially warmer area.
Unplug the device from the power supply and allow it to acclimatize for about 2
hours at room temperature.
3.3
Leveling the equipment
Purpose:
-
to compensate for unevenness at the place of installation
-
to ensure that the device is placed in a perfectly horizontal position for consistently
reproducible weighing results
Always re-level the device any time after it has been moved to a different location.
.
Level the scale using the four leveling feet.
.
Ensure that all leveling feet make contact with the floor.
.
Turn the feet until the air bubble is centered in the level indicator.
Each of the leveling feet must support an equal load.