background image

The scale will then display [SET.HI] (set high limit).

11. Press the

Unit

button to increase or the

Mode

F

button to decrease the digit value.

12. Press the

Off

(Yes) button to accept and highlight the next digit.

The 2nd digit is blinking [0_0000].

13. Press the

Unit

button to increase or the

Mode

F

button to decrease the digit value.

14. Repeat until all the digits are correct.

15. Press the

Off

(Yes) button to confirm.

The Display will show the low limit value with all digit blinking [000500] (in this
example, 500 pcs).

16. Press the

Off

(Yes) button accept the low limit value.

17. Same procedure like low limit setting.

The Display will show the high limit value with all digit flashing [00510.0] (in this
example, 510 pcs).

18. Press the

Off

(Yes) button to accept high limit value.

The scale is ready for check counting application with new checklimit. Go on step
21.

5 Operation

Bench scale Puro®

Minebea Intec

EN-23

All manuals and user guides at all-guides.com

Summary of Contents for Bench scale Puro 150

Page 1: ...1 0 0 03 05 2020 Checked on the basis of Puro scale with software version 2 89 Minebea Intec Bovenden GmbH Co KG Leinetal 2 37120 Bovenden Germany Phone 49 551 309 83 0 Fax 49 551 309 83 190 All manua...

Page 2: ...spondence concerning this product the type name and release number serial number as well as all license numbers relating to the product have to be cited Note This document is partially protected by co...

Page 3: ...Battery Power 6 3 4 Connections 7 3 4 1 Connect printer 7 3 5 Mounting bracket 8 4 Device description 9 4 1 Operating 9 4 1 1 Display and operating elements 9 5 Operation 12 5 1 Basic Weighing Functio...

Page 4: ...alibration Procedure 40 7 3 GEO Geographical Factor Adjustment Procedure 43 7 4 GEO Code Table 45 8 SBI Interface 47 9 Maintenance repairs cleaning 49 9 1 Repairs 49 9 2 Cleaning 49 9 2 1 Cleaning ins...

Page 5: ...for preventing hazards are not followed DANGER Warning of personal injury DANGER indicates death or severe irreversible personal injury which will occur if the corresponding safety measures are not ob...

Page 6: ...operty and or the environment may occur if appropriate safety measures are not observed Take the corresponding safety precautions Note User tips useful information and notes Bench scale Puro 1 Introdu...

Page 7: ...servicing or making internal connections The housing may only be opened by authorized and qualified personnel Do not use loads that exceed the capacity of the scale 2 2 Incoming goods inspection The s...

Page 8: ...e device to extreme moisture over long periods 3 3 Power supply AC power is used to power the scale when battery power is not needed Plug the USB C plug into the USB C jack on the bottom of the unit t...

Page 9: ...scale must be performed in a dry environment WARNING Risk of explosion can occur If the rechargeable battery is replaced with the wrong type or if it is not properly connected Battery is to be replace...

Page 10: ...et over the threaded holes in the side of the indicator and install the knobs Adjust the indicator to the desired angle and tighten the knobs Bench scale Puro 3 Device installation EN 8 Minebea Intec...

Page 11: ...ents LCD Display 1 2 16 15 14 3 4 5 6 7 8 9 10 11 12 18 17 13 Item Description Item Description 1 Busy 10 Low Average Piece Weight 2 Plus symbol 11 Low Sample Weight 3 Minus symbol 12 Data transfer 4...

Page 12: ...ing rules High OK Low Red Weight value upper tolerance limit Green Weight value is within OK range Yellow Weight value lower tolerance limit 4 1 1 3 Operating elements Button Off T Menu M Mode F Unit...

Page 13: ...o short press 1 second Accepts the current setting on the display Exits the user menu Aborts the cali bration in pro gress Advances to the previous digit in input process Moves back to previous menu i...

Page 14: ...starts with these parameters If the scale is turned ON for the very first time then applications Weighing plus Totalizing manually are active All elements of the display are shown for 2 seconds Press...

Page 15: ...pplications Totalization Auto Tare and Automatic Print can be activated via Menu OP FUNC A TARE Auto Tare TOT SET Totalization PRINT A PRINT Automatic Print 1 To select the weighing mode press and hol...

Page 16: ...splay of a weight value between different weight units Possible units are gram g kilogram kg pound lb ounce oz pound ounces lb oz For taring the container on the scale press the T button if the pan is...

Page 17: ...re is already a piece weight initialized CLR PW is displayed Otherwise PUT 20 for example is displayed then go on at step 4 3 Press the Unit No button to use the stored reference weight 4 Press the Of...

Page 18: ...2 g weight on the pan 2 0 g then LOW REF is displayed But is the weight on the pan 1 8 g then REF ERR is displayed 9 Press the Off Yes button to store the reference quantity sample The display shows...

Page 19: ...tion of the item until it is within the ACCEPT range Zero Check For comparison of subsequent samples with an initial reference samp le In this case the UNDER limit must be negative value and the OVER...

Page 20: ...at step 11 3 Press the Off Yes button to set new limits SET LOW set low limit is displayed and the yellow LED for the low limit is on 4 Press the Off Yes button to view the Low limit value The Display...

Page 21: ...this example 1 000 kg 8 Repeat until all the digits are correct 9 Press the Off Yes button to accept the low limit value when all digits are blinking again The scale will then display SET HI set high...

Page 22: ...use stored check limit value 14 Place container on the scale in this example 0 527 kg 15 Press the T button with NET appear beside the weight value 16 Place sample in the container 17 To view the curr...

Page 23: ...e Mode F button until CHECK is displayed with application icons and at bottom of the screen 2 Press the Unit No button to use the stored piece weight PW and go on at step 11 3 Press the Off Yes button...

Page 24: ...the display shows REF ERR reference error briefly 7 Press the Unit No button to use the stored limits and go on at step 18 8 Press the Off Yes button to set new limits The scale will then display SET...

Page 25: ...15 Press the Off Yes button to confirm The Display will show the low limit value with all digit blinking 000500 in this example 500 pcs 16 Press the Off Yes button accept the low limit value 17 Same p...

Page 26: ...0 Press the Off Yes button to establish new check limit value or press the Unit No button to use stored check limit value 21 Place container on the scale in this example 109 pcs 22 Press the T button...

Page 27: ...Totalization manual is set to ON by default 5 2 4 1 Procedure on Setting up the Totalization Application Release the button 24 To view the current check limit reference value In this example low limi...

Page 28: ...ub menu item 5 Change menu item until TOT SET totalizing setting is shown on display 6 Press the Off Yes button to enter the displayed sub menu then select one option OFF AUTO MAN using the Unit butto...

Page 29: ...e next item can be accumulated 5 2 4 3 Viewing and Clearing Statistical Data The pan must be empty to view the stored statistics 3 Empty the weighing pan 4 Put another sample on scale and follow same...

Page 30: ...nimum value Min 0 470 kg Maximum value Max 1 485 kg To clear the totalizing memory 2 Press the Menu M key to display the stored totalizing data 3 Press and hold the T button while the pan is unloaded...

Page 31: ...annot be added to the same total If the first load is a gross weight following loads must also be gross weights If the first load is a net weight following loads must also be net weights The message C...

Page 32: ...ton until M E N U is displayed The 1st menu item APPLIC shown on display 2 To enter into menu item in this example APPLIC WEIGH press Off Yes button 3 Or press the Unit button to move to the next menu...

Page 33: ...hapter 6 2 11 E N D Leave menu WEIGH Weighing mode OFF disabled ON enabled COUNT Count mode OFF disabled ON enabled CHECK Check mode 6 2 Menu Navigation Overview of Menu Mode options 6 2 1 Menu Select...

Page 34: ...off after 10 minute of no activities DYN TIM has no effect in this device 5 SEC has no effect in this device 10 SEC has no effect in this device 15 SEC has no effect in this device 20 SEC has no effec...

Page 35: ...ER Alarm when the weight is under the low limit OVER Alarm when the weight is over the high limit UNDOVR Alarm when the weight is outside Accept range BEEP KE Button Sound OFF disabled ON enabled TOT...

Page 36: ...TER Printing at the defined interval 1 3600 1 3600 second CONT Print continously ACCEPT Print on stable and within checklimit CONTNT Content of a printout RESULT Display Value OFF disabled ON enabled...

Page 37: ...4800 9600 9600 19200 19200 38400 38400 57600 57600 115200 115200 PARITY Parity 7 EVEN 7 data bits even parity 7 Odd 7 data bits odd parity 7 NONE 7 data bits no parity 8 NONE 8 date bits no parity STO...

Page 38: ...7600 115200 115200 PARITY Parity 7 EVEN 7 data bits even parity 7 Odd 7 data bits odd parity 7 NONE 7 data bits no parity 8 NONE 8 date bits no parity 7 MARK 7 data bits no parity 7 SPACE 7 data bits...

Page 39: ...S METRO Metrology Menu OFF Unlock ON Locked S PRINT Print Menu OFF Unlock ON Locked S PR COM Communication Menu OFF Unlock ON Locked S PC OUT PC Output Menu OFF Unlock ON Locked S PC COM PC Communicat...

Page 40: ...re accurate weighing results Before performing the calibration be sure to have the appropriate calibration weights Adjust the GEO setting according to your location see table 7 4 CAL ADJ 7 1 CAL Calib...

Page 41: ...00 0 in this example 150 kg C is displayed while zero reading is stored 8 For changing the calibration weight value press the Unit No button to edit the value The first digit is blinking _015 000 kg 9...

Page 42: ...ts are correct The Display shows the calibration value with all digit blinking 00100 0 in this example 100 kg 12 When calibration weight value is correct put specified weight on pan 13 Press Off Yes b...

Page 43: ...nge menu item until CAL ADJ shows on display 4 Press Off Yes button enter into sub menu item 5 Change menu item until LIN shows on display 6 Press Off Yes button to start zero point calibration 0 kg s...

Page 44: ...ity adjustment at 50 of scale capacity The display shows the the 2nd linearization weight value at 100 of scale capacity with all digit blinking 00150 0 in this example 150 kg C will be displayed whil...

Page 45: ...to table 7 4 to determine the GEO factor that corresponds to your location Enter into Menu Mode Release the button The display shows the linearization weight value at 100 of scale capacity 00150 0 in...

Page 46: ...ted and blinking 8 If changing the value is needed selection from 0 31 press the Unit button to increase or the Mode F button to decrease the GEO factor value 9 Press Off Yes button to confirm GEO Fac...

Page 47: ...25 21 26 45 10 9 9 8 8 7 7 6 6 5 5 26 45 28 06 10 10 9 9 8 8 7 7 6 6 5 28 06 29 25 11 10 10 9 9 8 8 7 7 6 6 29 25 30 41 11 11 10 10 9 9 8 8 7 7 6 30 41 31 56 12 11 11 10 10 9 9 8 8 7 7 31 56 33 09 12...

Page 48: ...52 57 04 23 22 22 21 21 20 20 19 19 18 18 57 04 58 17 23 23 22 22 21 21 20 20 19 19 18 58 17 59 32 24 23 23 22 22 21 21 20 20 19 19 58 17 59 32 24 23 23 22 22 21 21 20 20 19 19 60 49 62 90 25 24 24 2...

Page 49: ...ossible characters 0 9 E Unit possible characters a z A Z CR Carriage Return ASCII 0x0D LF Line Feed ASCII 0x0A Zero the weighing platform ESC Z EOC Answer see special Response Commands Tare the weigh...

Page 50: ...eported an Error during command execution LOCKED 1 2 3 4 5 L O C CR LF The command can not be executed because the parameter is currently locked Bench scale Puro 8 SBI Interface EN 48 Minebea Intec Al...

Page 51: ...NOTICE Some cleaning agents may not be compatible with the device material Do not use cleaning agents for stainless steel parts that contain chlorine alkalines acetic acid hydrochloric acid sulfuric...

Page 52: ...n by local public waste disposal companies EU legislation requires its Member States to collect electrical and electronic equipment and dispose of it separately from other unsorted municipal waste so...

Page 53: ...apter does not resolve or describe your problem contact your authorized service agent Please visit our website http www puroscales com to locate the office nearest you 11 Troubleshooting Bench scale P...

Page 54: ...izing Construction material in dicator P Plastic P painted carbon steel frame 1 2 mm 304 stanless steel load plate Construction material platform S all 304 stainless steel structure Indicator Weighing...

Page 55: ...ator and adjustable non slip levelling feet Feet 4 adjustable non slip levelling feet Load cell protection IP67 Load cell excitation volta ge maximum 18V AC DC recommended 5V 15V AC DC DD 753 x 596 x...

Page 56: ...12 4 Dimensions EF 4P Bench scale Puro 12 Technical data EN 54 Minebea Intec All manuals and user guides at all guides com a l l g u i d e s c o m...

Page 57: ...all dimensions in mm inch all dimensions in mm inch all dimensions in mm inch 12 Technical data Bench scale Puro Minebea Intec EN 55 All manuals and user guides at all guides com...

Page 58: ...all dimensions in mm inch Bench scale Puro 12 Technical data EN 56 Minebea Intec All manuals and user guides at all guides com...

Page 59: ...EF 4PDDP EF 4PDDS all dimensions in mm EF 4PFEP EF 4PFES all dimensions in mm 12 Technical data Bench scale Puro Minebea Intec EN 57 All manuals and user guides at all guides com...

Page 60: ...EF 4PGFP EF 4PGFS all dimensions in mm Bench scale Puro 12 Technical data EN 58 Minebea Intec All manuals and user guides at all guides com...

Page 61: ...is ON If Printx Content Info is ON Weigh Mode with Totalization Printout Description Comment 11 11 kg N 1 23 kg T 11 11 kg N 12 34 kg G MODE WEIGHT N 4 TOTAL 50 35 kg MIN 11 11 kg MAX 14 85 kg Result...

Page 62: ...uarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipme...

Page 63: ...Published by Minebea Intec Bovenden GmbH Co KG Leinetal 2 37120 Bovenden Germany Phone 49 551 309 83 0 Email info minebea intec com www minebea intec com All manuals and user guides at all guides com...

Reviews: