The display shows the 1
st
linearization weight value (50% of scale capacity)with all
digit blinking [00075.0] (in this example, 75 kg). This value cannot be modified.
[--C--] will be displayed while zero reading is stored.
9. Put the 1
st
linearization weight on the pan.
10. Press
Off
(Yes) button to start linearity adjustment at 50% of scale capacity
The display shows the the 2
nd
linearization weight value at 100% of scale capacity
with all digit blinking [00150.0] (in this example, 150 kg).
[--C--] will be displayed while reading is stored.
11. Put 2
nd
linearization weight on the pan.
12. Press
Off
(Yes) button to start linearity adjustment at 100% of scale capacity.
[--C--] is displayed while reading is stored.
Bench scale Puro®
7 Calibration and Adjustment
EN-42
Minebea Intec
All manuals and user guides at all-guides.com