BeneVision N22/BeneVision N19 Patient Monitor Service Manual
6-15
6.5.9
Removing Battery Backplane
As shown in the following picture, loosen and remove the four M3X6 cross recessed pan head screws with pad, then
remove the battery cavity assembly; remove the two M3X6 cross recessed pan head screws with pad, then remove the
battery backplane.
6.6
Disassembling the Module Rack
6.6.1
Disasembling the Handle and Hooks
As shown in the following figure, loosen and remove the eight M3X10 cross recessed countersunk head screws, and
remove the hooks.
As shown in the following figure, loosen and remove the three M3×8 cross recessed countersunk head screws, and
remove the handle.
M3 screw
M3 screw
M3X10 screw
M3X8 screw