2-12
BeneVision N22/BeneVision N19 Patient Monitor Service Manual
2.4.4
Module Initialization
Parameter module power-on sequence:
Power-on description using MPM, N1, or T1 as an example:
Insert MPM into SMR
The module rack uploads the module
availability status
The system identifies the status and sends
the handshake command
Query for 1 minute continuously until
success is achieved.
Check status self test
Status self test succeeded
Parameter configuration
Configuration succeeded
The device operates properly
Module failure
Module failure
Module failure