BeneVision N22/BeneVision N19 Patient Monitor Service Manual
A-3
A.6
Electrical Safety Inspection Test
Inspection and Testing
Limit
1
Power Cord Plug
2
Device Enclosure and Accessories
/
3
Device Labeling
/
4
Protective Earth Resistance
Max 0.2 Ω
5
Earth Leakage
Normal condition(NC)
Max:
NC: 300μA(refer to UL60601-1)
SFC: 1000μA
Single Fault condition(SFC)
6
Patient Leakage Current
Normal condition(NC)
Max:
CF applied part:
NC:10μA, SFC: 50μA
BF applied part:
NC:100μA, SFC: 500μA
Single Fault condition(SFC)
7
Mains on Applied Part Leakage
Max:
CF applied part: 50μA
BF applied part: 5000μA
8
Patient Auxiliary Current Normal condition(NC)
Max:
CF applied part:
NC:10μA, SFC: 50μA
BF applied part:
NC:100μA, SFC: 500μA