BeneVision N22/BeneVision N19 Patient Monitor Service Manual
6-1
6
Disassembly and Repair
6.1
Tools
During disassembly and repair, the following tools may be required:
Phillips screwdrivers
Tweezers
Needle nose pliers
Cutting pliers
Flat-bladed screwdriver
6.2
Preparations for Disassembly
Before disassembling the monitor, make following preparations:
Stop monitoring the patient, turn off the monitor and disconnect all the accessories and peripheral devices.
Disconnect the AC power supply then remove the battery. Before taking out the battery, remove the main unit
housing.
6.3
Whole Unit Disassembly
NOTE
Before disassembly, make sure that the work surface is clean, smooth and free of debris that could cause
damage to the touchscreen.
All disassembly and repairs should only be performed by qualified service personnel using anti static
precautions.