Instruction Manual Mi 180 Bench Meter
4 0
4 0
4 0
4 0
4 0
ELECTRODE CONDITIONING & MAINTENANCE
PREPARATION PROCEDURE
PREPARATION PROCEDURE
PREPARATION PROCEDURE
PREPARATION PROCEDURE
PREPARATION PROCEDURE
Remove the electrode protective cap.
DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT. This is normal with
electrodes and they will disappear when rinsed with water.
During transport tiny bubbles of air may have formed inside the glass bulb. The electrode
cannot function properly under these conditions. These bubbles can be removed by “shaking
down” the electrode as you would do with a glass thermometer.
If the bulb and/or junction are dry, soak the electrode in MA 9015
MA 9015
MA 9015
MA 9015
MA 9015 storage solution for
at least one hour.
For refillable electrodes, if the refill solution (electrolyte) is more than 2½ cm (1”) below
the fill hole, add the appropriate electrolyte solution.
M E A S U R E M E N T
M E A S U R E M E N T
M E A S U R E M E N T
M E A S U R E M E N T
M E A S U R E M E N T
Rinse the electrode tip with distilled water, immerse it (4 cm / 1½") into the sample and stir
gently for a few seconds.
For a faster response and to avoid cross contamination of the samples, rinse the electrode
tip with the solution to be tested before taking any measurements.
STORAGE PROCEDURE
STORAGE PROCEDURE
STORAGE PROCEDURE
STORAGE PROCEDURE
STORAGE PROCEDURE
To minimize clogging and ensure a quick response time, the glass bulb and the junction
should always be kept moist.
When not in use, store it with a few drops of M
M
M
M
MA
A
A
A
A 9015
9015
9015
9015
9015 storage solution in the protective cap.
NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED WATER.
LAMV
LAMV
LAMV
LAMV
LAMV
Requests all mV/Rel mV Log on demand.
L A E C
L A E C
L A E C
L A E C
L A E C
Requests all EC Log on demand.
LATD
LATD
LATD
LATD
LATD
Requests all TDS Log on demand.
L A N C
L A N C
L A N C
L A N C
L A N C
Requests all NaCl Log on demand.
Notes
Notes
Notes
Notes
Notes:
•“Err8” is sent if the instrument is not in measurement mode.
•“Err6” is sent if the requested range is not available.
•“Err4” is sent if the requested set parameter is not available.
•“Err3” is sent if the Log on demand is empty.
•Invalid commands will be ignored.