•When the reading is out of range, the closest full-scale value is displayed blinking.
•If using a pH electrode while in mV mode, the instrument will measure the mV generated
by the pH electrode.
•If the instrument displays a Rel mV reading and it is desired to take mV measurements,
simply clear the Rel mV calibration (see Rel mV calibration section at page 16).
CONDUCTIVITY
CONDUCTIVITY
CONDUCTIVITY
CONDUCTIVITY
CONDUCTIVITY MEASUREMENTS
MEASUREMENTS
MEASUREMENTS
MEASUREMENTS
MEASUREMENTS
Make sure the instrument has been calibrated before taking conductivity measurements.
• Press the RANGE/FIXED key to enter EC measurement mode.
• Immerse the probe into the solution to be tested. The sleeve holes must be completely submerged.
Tap the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve.
• The conductivity value is displayed on the primary LCD and the temperature on the secondary
LCD, along with the reference temperature.
Notes
Notes
Notes
Notes
Notes:
•If the reading is out of range, the closest full-scale value (200.0 mS for MTC/ATC mode or
500.0 mS for uncompensated conductivity) will be displayed blinking.
•If SHIFT&RANGE/FIXED keys are pressed to freeze the LCD range and the reading goes
out of range, the full-scale value of the frozen range will be displayed blinking.
The EC reading is affected by temperature. Three options for temperature compensation are
available in EC measurement mode.
or