PERIODIC MAINTENANCE
PERIODIC MAINTENANCE
PERIODIC MAINTENANCE
PERIODIC MAINTENANCE
PERIODIC MAINTENANCE
Inspect electrode and cable. The cable used for the connection to the instrument must
be intact and there must be no points of broken insulation on the cable or cracks on the
electrode stem or bulb. If any scratches or cracks are present, replace the electrode.
Rinse off any salt deposits with water.
Connectors must be perfectly clean and dry.
For refillable electrodes
:
Refill the electrode with fresh electrolyte (see the electrode’s specifications to select the
correct refilling solution). Allow the electrode to stand upright for 1 hour. Follow the Storage
Procedure above.
pH CLEANING PROCEDURE
pH CLEANING PROCEDURE
pH CLEANING PROCEDURE
pH CLEANING PROCEDURE
pH CLEANING PROCEDURE
•
General
Soak in MA
MA
MA
MA
MA 9016
9016
9016
9016
9016 General Cleaning Solution for approx. ½ hour.
IMPORTANT: After performing any of the cleaning procedures, rinse the electrode thoroughly
with distilled water and soak it in M
M
M
M
MA
A
A
A
A 9015
9015
9015
9015
9015 storage solution for at least 1 hour before taking
measurements.
EC PROBE MAINTENANCE
EC PROBE MAINTENANCE
EC PROBE MAINTENANCE
EC PROBE MAINTENANCE
EC PROBE MAINTENANCE
Rinse the probe with clean water after measurements. If a more thorough cleaning is
required, remove the probe sleeve and clean the probe with a cloth or a nonabrasive
detergent. Make sure to reinsert the sleeve onto the probe properly and in the right
direction. After cleaning the probe, recalibrate the instrument.
Take great care while handling the probe.