1. Scope 

  This service manual applies to 2266-20 Laser Temp-Gun™ Thermometer.

1.0 

EQUIPMENT USE: 

1.1 Below is Recommended Reference Thermometer Traceable to NIST: 

Manufacturer 

Model No. 

Range 

Accuracy 

Hart Scientific 

1502A 

-200°C to 962°C 

±O.OO6°C accuracy at 0°C 

1.2 Below are recommended IR Black Bodies Sources 

/

TC Calibrator: 

Manufacturer 

Model No. 

Range 

Accuracy 

Emissivity 

Target Size 

Hart Scientific 

9132 

50°C~ +500°C 

±0.5°C accuracy at 

100°C 

0.95 

(±0.02) 

2.25in 

(57mm) 

Hart Scientific 

9133 

-35°C~ +150°C 

±0.0°C accuracy at 

100°C 

0.95 

(±0.02) 

2.0in 

(57mm) 

Hart Scientific 

4181 

35°C~ +500°C 

±0.5°C accuracy at 

100°C 

0.90~0.95 

(±0.02) 

6.0in 

(152mm) 

Fluke TC 

724/714 

-200°C~ +1370°C

±0.5°C accuracy at 

100°C 

NA

NA) 

1.3 Operate only the test equipments within its calibration period. 

2.0 

TESTING CONDITION AND INTERFERENCE: 

2.1  Calibration and verification are performed at an ambient temperature of 23°C - 25'C and ambient 

relative  humidity  40% 

+1- 

10%  in  a  controlled  room.  Each  thermometer  should  be  uniquely 

identified as to the room 

andlor 

area in which it is used or tested. 

2.2  If the meter is exposed to significant changes in ambient temperature (hot to cold or cold to hot) 

allow 30 minutes  for  the unit  to  acclimate  with room  temperature  before  taking  any  temperature 

reading or calibration. 

2.3  Operate the unit under test (UUT) and the reference thermometer within its specified range. 
2.4  Calibration can be verified using a certified NIST traceable standard thermometer.  Recalibration 

must be performed in similar environment conditions as it was calibrated to get similar results. 

2.5  Strong electromechanical radiation or static discharges can cause reading errors if situated near 

the meter. 

2.6  The  object  size  under  test  should  be  larger  than  the  spot  size  calculated  by  the  field  of  view 

diagram of the meter. 

3.0 

 FIELD CALIBRATION PROCEDURE: 

3.1

IR

Software Calibration Check using Black Body Source (Calibrator 9133 or 4181) 

3.1.1   Condition the unit to be calibrated in a stable environment as described in 3.1 for at least 

30 minutes. 

5

Summary of Contents for 2266-20 Laser TEMP-GUN

Page 1: ...erform all calibration at an ambient temperature of 23 C 2 C 73 4 F 3 6 F and relative humidity of 80 allow the TEMP GUN thermometer to sit at this temperature for a minimum of 30 minutes before proce...

Page 2: ...Specifications 3 Basic Maintenance 4 Loading Changing the Batteries 4 Maintaining Tool 4 Cleaning the Laser Windows 4 Scope 5 Equipment Use 5 Testing Condition and Interference 5 Field calibration Pr...

Page 3: ...nual If you do not do so the protection provided by the Laser TEMP Gun may be impaired Read the Safety Information page before servicing this product In this manual a WARNING identifies conditions and...

Page 4: ...ctices that could result in personal injury or loss of life Specific precautions To reduce the risk of injury user must read and understand operator s manual Laser light Do not stare into beam or view...

Page 5: ...2 C 4 F or 2 of reading whichever is greater Assume ambient operating temperature of 23 C to 25 C 73 F to 77 F Min measuring distance 2 50 C 122 F 4 50 C 122 F Display resolution 0 1 C 0 1 F Contact t...

Page 6: ...r operate at full power with fully charged batteries clean the contacts on the battery door If the tool still does not work properly return the tool to a MILWAUKEE service facility for repairs Cleanin...

Page 7: ...40 1 10 in a controlled room Each thermometer should be uniquely identified as to the room andlor area in which it is used or tested 2 2 If the meter is exposed to significant changes in ambient tempe...

Page 8: ...mometer as shown in Figure 1 if necessary 3 1 6 Position correctly the IR opening towards the black body target source at 3 inches apart for Fluke 9133 or 11 inches for Fluke4181 Ensure that the senso...

Page 9: ...ration mode 3 2 4 Press ALM button to skip 25 C as this is for factory calibration only 3 2 5 32 F or O C degrees will then appear for next calibration point 3 2 6 Set the Fluke 714 724 calibrator as...

Reviews: