MILWAUKEE ELECTRIC TOOL CORPORATION – TEST & MEASUREMENT PRODUCT 

3.2

KTC Software Calibration Check using Temperature Calibrator (Fluke 714/724) 

3.2.1  Condition the unit to be calibrated in a stable environment as described in 3.1 for at least 30 

minutes. 

3.2.2  Connect the KTC output of Fluke714/724 to the unit's KTC input. See Figure 3. 
3.2.3  Press and Hold the ALM button, then apply power. Press the Trigger button 4 times.  After 4 

seconds, the unit will enter into KTC calibration mode. 

3.2.4  Press ALM button to skip "25°C" as this is for factory calibration only. 
3.2.5  32°F or O°C degrees will then appear for next calibration point. 
3.2.6  Set the 

Fluke

714/724 calibrator as the source of K type output to be calibrated e.g. 0, 50, 

100, 250, 450, -20 deg.C. 

3.2.7  Press Trigger button to confirm. "PASS" will display. 
3.2.8  Press ALM button to select the next desired temp. calibration point. 
3.2.9  Repeat step 4.26-2.8 for 50, 100,250,450, -20 deg. remaining calibration points. 
3.2.10  You can skip any calibration points by pressing "ALM" button until the unit will exit and turn 

off. 

3.2.11  Once exit, the unit will default to Normal Temperature Mode. 

Figure 3 

7

Summary of Contents for 2266-20 Laser TEMP-GUN

Page 1: ...erform all calibration at an ambient temperature of 23 C 2 C 73 4 F 3 6 F and relative humidity of 80 allow the TEMP GUN thermometer to sit at this temperature for a minimum of 30 minutes before proce...

Page 2: ...Specifications 3 Basic Maintenance 4 Loading Changing the Batteries 4 Maintaining Tool 4 Cleaning the Laser Windows 4 Scope 5 Equipment Use 5 Testing Condition and Interference 5 Field calibration Pr...

Page 3: ...nual If you do not do so the protection provided by the Laser TEMP Gun may be impaired Read the Safety Information page before servicing this product In this manual a WARNING identifies conditions and...

Page 4: ...ctices that could result in personal injury or loss of life Specific precautions To reduce the risk of injury user must read and understand operator s manual Laser light Do not stare into beam or view...

Page 5: ...2 C 4 F or 2 of reading whichever is greater Assume ambient operating temperature of 23 C to 25 C 73 F to 77 F Min measuring distance 2 50 C 122 F 4 50 C 122 F Display resolution 0 1 C 0 1 F Contact t...

Page 6: ...r operate at full power with fully charged batteries clean the contacts on the battery door If the tool still does not work properly return the tool to a MILWAUKEE service facility for repairs Cleanin...

Page 7: ...40 1 10 in a controlled room Each thermometer should be uniquely identified as to the room andlor area in which it is used or tested 2 2 If the meter is exposed to significant changes in ambient tempe...

Page 8: ...mometer as shown in Figure 1 if necessary 3 1 6 Position correctly the IR opening towards the black body target source at 3 inches apart for Fluke 9133 or 11 inches for Fluke4181 Ensure that the senso...

Page 9: ...ration mode 3 2 4 Press ALM button to skip 25 C as this is for factory calibration only 3 2 5 32 F or O C degrees will then appear for next calibration point 3 2 6 Set the Fluke 714 724 calibrator as...

Reviews: