MILWAUKEE ELECTRIC TOOL CORPORATION – TEST & MEASUREMENT PRODUCT
3.2
KTC Software Calibration Check using Temperature Calibrator (Fluke 714/724)
3.2.1 Condition the unit to be calibrated in a stable environment as described in 3.1 for at least 30
minutes.
3.2.2 Connect the KTC output of Fluke714/724 to the unit's KTC input. See Figure 3.
3.2.3 Press and Hold the ALM button, then apply power. Press the Trigger button 4 times. After 4
seconds, the unit will enter into KTC calibration mode.
3.2.4 Press ALM button to skip "25°C" as this is for factory calibration only.
3.2.5 32°F or O°C degrees will then appear for next calibration point.
3.2.6 Set the
Fluke
714/724 calibrator as the source of K type output to be calibrated e.g. 0, 50,
100, 250, 450, -20 deg.C.
3.2.7 Press Trigger button to confirm. "PASS" will display.
3.2.8 Press ALM button to select the next desired temp. calibration point.
3.2.9 Repeat step 4.26-2.8 for 50, 100,250,450, -20 deg. remaining calibration points.
3.2.10 You can skip any calibration points by pressing "ALM" button until the unit will exit and turn
off.
3.2.11 Once exit, the unit will default to Normal Temperature Mode.
Figure 3