TYPICAL H-FLUME
point of measurement
transducer
D
plan
front
side
Q
For rated flows under free flow conditions, the head is measured at a point downstream from the flume
entrance. Refer to the following table.
Flume Size
Point of Measurement
D (feet)
cm
inches
0.5
4.7
1.88
0.75
6.7
2.69
1.0
9.1
3.63
1.5
13.5
5.38
2.0
17.9
7.19
2.5
22.5
9.00
3.0
27.2
10.88
4.5
40.5
16.19
Position the transducer such that it is centered over the flow at a minimum height of 60 cm (2ft) above the
maximum head.
PL-269
7 – 36