SECTION VI
AUXILIARY OPERATIONS
F1
EMULATION MODE
The flow calculation parameters can be checked for accuracy by using the emulation function, F1. A
liquid head is entered via the keypad and the OCM II will respond by displaying the flow associated with
that head. During emulation mode, all totalizing stops. This feature may be used to find a head that is
equal to a flowrate that will suit the chart recorder better.
e.g.
If the maximum flowrate of a system is 721.6 litres / second, however the desired maximum
flowrate is 700 litres / second the operator may key in head values to find the head value
required to produce the desired flowrate.
Enter: [F0] [E] (if not yet in the access mode)
[2.71828] [E]
"P0" will appear
Enter:
[F1] [E]
[HEAD
1
] [E]
"FLOW" corresponding to HEAD
1
will be displayed.
Enter:
[HEAD
2
] [E]
"FLOW" corresponding to HEAD
2
will be displayed, etc...
F3
CLEARING TOTALIZERS
Used to clear all totalizers. This will zero displays: D2, accumulated flow; D10-D21, % of accumulated
flow and D22, serial port totalizer
Enter:
[F0] [E] (if not yet in access mode)
[2.71828] [E]
"P0" will appear
Enter:
[F3] [E]
[2.71828] [E]
"clr" will appear, indicating that the totalizers have been cleared.
PL-269
6 – 1
Summary of Contents for OCM II PL-269
Page 6: ......
Page 8: ......
Page 12: ......
Page 22: ......
Page 28: ......
Page 29: ......
Page 30: ......
Page 31: ......
Page 32: ......
Page 33: ......
Page 34: ......
Page 35: ......
Page 36: ......
Page 76: ......
Page 78: ......
Page 92: ...PL 269 9 14 ...
Page 94: ......
Page 95: ......
Page 106: ......