Page 14
Milltronics Weight Lifter MWL – Instruction Manual
7ML19981CR01
7. Test the loaded MWL
• Keep the lifting-pads raised, and add the other calibration weights
• Check again to ensure there is 10mm (0.38") minimum clearance between the
side of the stacked weights and the vertical arms of the calibration-weight
supports. If necessary, move the two crank bodies further from the calibration-
weight supports to achieve this clearance.
• Recheck the crank body alignment: (See “Align the crank bodies to each other, if
necessary” on page 10)
Lifting pads raised
Calibration weights
applied
Calibration-weight support (2)
Calibration weights
Vertical arm of calibration-
weight support
Minimum clearance on MUS scales: 10mm (0.38")
on MSI Scales: 12.5mm (0.5")
MWL48all.fm Page 14 Tuesday, June 12, 2001 8:23 AM