80
Replacing the Q-Pak® Pack,
Continued
Placing
(continued)
Step
Action
Diagram
3
Push the Pack Locking Handle
down.
Close the left door.
4
When a new Q-Pak® Pack is
installed, the screen looks like
this.
Press
.
Rinsing
The Q-Pak® pack, when newly installed, needs to be rinsed. This ensures optimal water
quality.
Step
Action
Diagram
1
Locate the clear tubing and the
barbed fitting from the System
accessories bag.
Screw the barbed fitting onto
the POD Unit.
NOTE:
Do not use any white tape on
the threads of the barbed fitting.
An O-ring is located inside the
POD Unit.
Push one end of the clear tubing
onto the end of the barbed
fitting.
Place the other end of the clear
tubing into a sink.
Continued on next page
Summary of Contents for Direct 16