113
Summary of Alert messages,
Continued
Alert messages
(continued)
(continued)
TAP FEED CONDUCTIVITY > SP
•
The Tap Water conductivity is below
the set point.
•
Contact Millipore SAS.
THE NETWORK CABLE IS
UNPLUGGED
•
Check the Ethernet Cable plugged
into the System and the computer.
•
Restart the System.
THIS IP ADDRESS IS ALREADY
USED BY ANOTHER SYSTEM
•
Contact your network administrator.
•
Restart the System.
TOC < 1 PPB
•
Invalid TOC measurement.
•
Perform a TOC Curve Check.
•
Contact Millipore SAS.
TOC FEEDWATER < 15.3 M
Ω
.cm
•
Due to low resistivity at the inlet of
the UV Lamp, the TOC measurement
can not be performed properly
anymore.
•
Please replace the Q-Pak® Pack to
ensure valid TOC indications.