74
Maintenance Schedule
Consumables
Item
Maintenance needed
When
Progard® Cartridge
Replacement
Prompted to by an LCD
message.
Q-Pak® Pack
Replacement
POD Pak
Replacement
Prompted to by an LCD
message or as necessary.
UV Lamps
Item
Maintenance needed
When
UV 254 nm Lamp
Replacement
Prompted to by an LCD
message.
UV 185 nm Lamp
NOTE:
It is recommended to have a Millipore SAS Service Representative change the UV Lamps
in the system.
The replacement of this lamp involves removing the cover of the system. The
instructions for replacing these lamps are not included in this User Manual. The
instructions are included with the replacement lamp.
Cleaning/
Sanitisation
Item
Maintenance needed
When
Inlet Strainer
Cleaning
Prompted to by an LCD message
or as necessary.
RO
Cartridge(s)
Cl
2
cleaning
When prompted to by an LCD
message.
pH Cleaning
As necessary.
System
Sanitisation
Contact Millipore SAS for more
details.
Calibrating the
flow rate
Item
Maintenance needed
When
Flow rate
Recalibration
New Consumable, Sensor or
change to Feed water.
See ‘Calibrating the flow rate’ for
more information.
TOC Curve
Check
Item
Maintenance needed
When
TOC Indicator Update TOC Curve Check New Q-Pak® Pack installed, or
when prompted to by an LCD
message.
Summary of Contents for Direct 16
Page 1: ...User Manual Milli Q Direct 8 16 System...
Page 2: ......
Page 6: ......
Page 47: ...Milli Q Direct 8 16 System 47 Software Map Map...
Page 115: ......
Page 116: ...Milli Q Direct 8 16 System FTPF11486 V3 0 03 2013...