Calibrating the Flow rate,
Continued
Procedure
(continued)
Step
Action
Diagram
6
Place a 1 L Graduated Cylinder
under the POD Unit.
Press
.
7
Press
.
8
Water dispenses automatically
from the POD Unit.
Wait until it stops dispensing
water.
9
Measure the amount of water
(in ml) that was dispensed.
Suppose 870 ml was collected.
Input this using the Keypad.
10
Perform the flow calibration
again to improve accuracy.
Press
.
11
Press 3 times on
.