Program guide
169
– General program for analysis and water analysis and for water based cultures with
acid-soluble metallic salts such as Ca
2+
and Mg
2+
etc.
– for light to medium levels of soiling,
– for normal to high wash result requirements.
Conductivity limit (Option)
Water intake
Water drainage
Set [µS/cm]
Set [µS/cm]
Number of repeats
0 /
1
Number of repeats
0 /
1
Cleaning
Rinse
Final rinse
1
2
1
2
3
4
1
2
CW50
HW
HW
DI
DI
DI
DOS 3
DOS 1
DOS 3
0.3
0.4
0.1
50°C
75°C
70°C
2
3
2
1
1
1
On
= Customizable parameters
CW = cold water
HW = hot water
CWxx = CW proportion in mixed water as percentage (CW70 = 70 % CW + 30 % HW)
DI = aqua destillata, fully demineralised water, demineralized water
Min = Holding time in minutes
DOS 1 = cleaning detergent
DOS 3 = neutralizing agent
DOS 4 = DOS module