Programme
When to use
Cleaning agent
(when not using liquid
cleaning agent)
Follow manufacturer’s
instructions
SPECIAL AN
93°C-10’
For cleaning and thermal disinfection at 93°C with a
10-minute holding time in accordance with §18 IfSG
(German Institute for the prevention of the spread of
epidemics). Area of effectiveness A/B (fungicidal,
bactericidal and virucidal disinfection) for anaesthetic
instruments.
45 grammes in the clean-
ing agent dispenser
SPECIAL
93°C-10’
For cleaning and thermal disinfection at 93°C with a
10-minute holding time. Area of effectiveness A/B
(fungicidal, bactericidal and virucidal disinfection)
according to §18 IfSG (German Institute for the pre-
vention of the spread of epidemics).
30 grammes on the right
hand side door interior
varioTD
For general hygiene requirements, particularly suit-
able for sensitive and heavily blood soiled instru-
ments and for processing aluminium cassettes.
This programme meets the requirements of prEN ISO
15883-1 90 °C (+ 5 °C, - 0 °C) with a 5 min. holding
time.
Preferably a mildly alkaline
liquid cleaning agent
varioTD AN
For general hygiene requirements, particularly suit-
able for sensitive and heavily secretion soiled anaes-
thetic instruments.
This programme meets the requirements of prEN ISO
15883-1 80 °C (+ 5 °C, - 0 °C) with a 10 min. holding
time.
Preferably a mildly alkaline
liquid cleaning agent