1
Introduction
This manual provides descriptions and operating instructions for the Midtronics Celltron
Essential
CTE-1000 and CTE-1500 stationary battery string analyzers. It helps you understand
the parts of the analyzer and how to use it to test batteries.
Safety instructions
Important
Read the instructions below before you operate the analyzer.
Guidelines
To avoid electric shock when testing batteries,
follow your company safety practices and these
guidelines:
•
Wear safety glasses or a face shield.
•
Wear protective rubber gloves.
•
Wear a protective apron or shop coat.
•
Perform service work only for which you have been trained.
•
Do not disconnect battery cables from power systems without authorization for the length
of time needed to complete testing.
•
Avoid placing yourself into a circuit.
•
Avoid contact with frame racks and adjacent hardware that may be grounded while in
contact with the battery.
About the analyzer
Terminology
The analyzer and manual use the term “jar,” an international term for “battery.” A string is a
series of jars connected together by straps to provide power as a whole.
Accessories
Table 1 lists the accessories that come with the CTE-1000.
Table 1. TEC-4500 accessories
Accessory
Part
Number
Description
DuraProbes
C087
Cables with 5-inch probes and 2-inch openings to attach to large
battery terminals