7
4. PRINT RESULTS option
The
PRINT RESULTS
option allows you to print test results for a string you tested.
5. VIEW RESULTS menu
The
VIEW RESULTS
option allows you to view test results for a string you tested. When you
select the string, the screen displays these options:
1. STRING SUMMARY
2. REVIEW DATA SET
3. MAIN MENU
Table 8 describes these options.
Table 8. VIEW RESULTS menu options
Option
Description
STRING
SUMMARY
Lists these values as a summary of the string you tested:
•
AVG. % —
Average percentage of the reference value.
•
AVG. SIEMENS: —
Average conductance value.
•
TOTAL JARS: —
Total number of jars you tested in the string.
•
LOW: —
Jar number with the lowest percentage of the reference value.
•
HIGH: —
Jar number with the highest percentage of the reference value.
•
STRING —
Average p
ercentage of the string as compared to the jar in the string
with the highest conductance value.
REVIEW DATA
SET
Lists these values for a jar or strap:
•
voltage
•
conductance value
•
number of the jar or strap
•
percent of the reference value
The analyzer keeps the test results for a string until you erase the data.
Helpful hint:
The
RETEST
button provides a shortcut to the
REVIEW DATA SET
screen.
6. EXPORT DATA option
The EXPORT DATA option allows you to download test results through the infrared (IR)
receiver to a PC.
7. CHANGE REF. option
The
CHANGE REF.
option allows you to change the reference value for a string without changing
other values in the
SET PARAMETERS
menu. For more information about reference values, refer
to “Determining a reference value” in Chapter 2.