37
Probe tip is bent or stops retracting
Replacing a probe tip
To replace a probe tip:
1. Use pliers to grasp the probe tip at the top of the sleeve that encases it.
Caution:
Do not grasp the sleeve. You can damage the probe.
2. Pull the tip straight out.
3. Grasp the replacement tip with the pliers and insert it into the sleeve.
4. Push the probe tip into a soft surface, such as cardboard, until the tip reaches the bottom
of the probe sleeve.
Note:
To obtain replacement tips, contact Midtronics Customer Service and ask for the part
number C069 for the DuraProbes or C059 for the MiniProbes.