ScanMaker 1000XL Plus User Guide 31
6
5
6
5
4
4
3
3
2
1
2
1
0
NOTE: To scan the target
with the TMA 1600-III, put the
Black Plate on top of the target
to cover the target and shield
unwanted light.
Positioning the Target
Proper positioning of the target on the scanner is important for successful calibration.
Incorrect positioning of the target will result in an unsuccessful calibration.
There are a wide selection of calibration targets in the market; thus, we only list two
examples to demonstrate how to position IT8 target on the scanner.
A. Positioning the Reflective Target
(LaserSoft IT8, size 10 x 15-cm)
1. Place the target face down on the scanner glass surface, making sure that the
target is positioned correct side up toward the front of the scanner.
2. Using the U-shaped ruler as a point of reference, adjust the target so that it is
positioned at the center of the scanner glass bed.
6
5
6
5
4
4
3
3
2
1
2
1
0
Important
: Align the
template's front corners
firmly against the top
corners (left and right)
of the U-shaped ruler,
located at the front of
the scanner.
Make sure the woman's
portrait lies towards the right
side of the scanner
Keep calibration strip clear
and free of any obstruction
B. Positioning the Transparent Target
(Kodak IT8, size 4"x5")
1. Load the target face down inside the slot of the 4" x 5" Film Template.
2. Place the 4" x 5" Film Template with the loaded target on the scanner glass
surface, and orient the template with the “Microtek” logo facing up.