DOC 6501_Release V
ATS-6501 Users Guide
35
[OK] 2013-08-05T16:49:56Z
To verify the card is working properly use the
status hardware:outputs:slots:N
command (N=Slot
#) to obtain the current status for a particular card.
ATS-6501>status hardware:outputs:slots:4
[ports]
[1]
[status] good
[2]
[status] good
[3]
[status] good
[4]
[status] good
[power] enabled
[type] 4393A
[OK] 2013-08-05T16:53:04Z
2.4.2.13 4383A IRIG Input Card
This module has four input ports that can be used to recover time from an IRIG time source.
Ports 1 and 2 support DC IRIG (Default B007) and Ports 3 and 4 support AM IRIG (Default
B127). By default Port #1 is active, to select another source or IRIG format use the
irig_input
command. If the IRIG Source does not provide the year as part of the time code message (i.e.
B000/B120) the user can specify the year with the command.
ATS-6501>irig_input [slot#] [port#] [irig type] [year]
To verify the system settings use settings hardware:slots:N where N = the slot # where the card is
installed.
6501A_61>settings hardware:slots:6
[tsc4383]
[port]
[1]
[enabled] true
[format] B007
[2]
[enabled] false
[format] B007
[3]
[enabled] false
[format] B127
[4]
[enabled] false
[format] B127
To verify the card is working properly use the
status hardware:outputs:slots:N
command (N=Slot
#) to obtain the current status for a particular card. The selected IRIG Source will report as good,
the other ports will show as disabled.
6501A_61>status hardware:outputs:slots:6
[ports]
[1]